福禄克FLUKE
型号:5606
美国FullImmersionPRT
The 5606 Full Immersion PRTs are designed to perform in extreme environments where both the transition junction and the lead wires are required to withstand temperatures covering the entire operating range of the probe. (“Immersion” in this case simply means exposure to temperatures other than ambient and does not necessarily refer to the medium to which the PRT and its components are subjected.)
Applications where “full immersion” of probe, transition junction, and lead wire is required can include calibration or validation of sensors used in laboratory or bio freezers, walk-in refrigerators, autoclaves, ovens, stability test chambers, furnaces or incubators. Other applications might include IQ/ OQ/PQ qualification procedures, temperature mapping, or data logging in temperature controlled spaces or environmental chambers.
The 5606 is just two inches (50 mm) in length with a sheath diameter of 1/8 inch (3.1 mm). Since it can be fully immersed over its entire temperature range, you don’t have to worry about calculating minimum immersion depth— immerse the entire probe, transition junction, and lead wires, in either non-corrosive liquids or dry mediums ranging from –200 °C to 160 °C.
The 100 Ω precision sensing element (alpha = 0.00385) is specially constructed to prevent the ingress of moisture while achieving calibration accuracies as good as ± 0.05 °C. You get both accuracy and flexibility built into a small, durable package.
The lead wires of the 5606 are made of single-conductor enameled copper wire which prevents moisture from wicking to the sensing element as it would if a more common, stranded lead wire were used. The diameter of the lead wire is no greater than 0.2 mm, making it easy to run lead wires through a freezer door to your readout or transmitter with effectively zero energy loss. The 5606 has been proven in common heat transfer fluids such as silicone oil, mineral oil, ethanol, and even liquid nitrogen with no degradation of the transition junction seal or lead wire insulation.
If you need longer lead wire than the standard length, call one of our Application Specialists today for a free quote. They can also help you determine which readout and probe termination are best suited for your application or help you select the right calibration for your probe.
General Specifications | |||
Temperature range |
|
||
Nominal resistance at 0.01 °C |
|
||
Temperature coefficient |
|
||
Accuracy[1] |
|
||
Short-term repeatability[2] |
|
||
Drift[3] |
|
||
Hysteresis |
|
||
Sheath length |
|
||
Sheath diameter |
|
||
Sheath material |
|
||
Transition junction temperature range[4] |
|
||
Sensor length |
|
||
Sensor location |
|
||
Minimum insulation resistance |
|
||
Minimum immersion length |
|
||
Maximum immersion depth in liquid medium |
|
||
Response time[5] |
|
||
Self heating (in 0 °C bath) |
|
||
Lead-wire cable type |
|
||
Lead-wire length |
|
||
Lead-wire temperature range |
|
||
Calibration |
|
[1] Includes calibration and 100 hr drift (k = 2).
[2] Three thermal cycles from min to max temp, includes hysteresis, 95 % confidence (k = 2).
[3] After 100 hours at max temp, 95 % confidence (k = 2).
[4] Temperatures outside this range will cause irreparable damage.
[5] Per ASTM E 644.
Calibration uncertainty | ||
5606 | ||
Temperature | 1924 Calibration Uncertainty† | Tolerance w/out Calibration‡ |
-197 °C | 0.03 °C | 0.4 °C |
-38 °C | 0.03 °C | 0.2 °C |
0 °C | 0.03 °C | 0.1 °C |
100 °C | 0.045 °C | 0.3 °C |
157 °C | 0.05 °C | 0.4 °C |
† Lab code 200348-0 ‡ Each PRT includes a certificate with a measured R0 value. An uncalibrated 5606 meets ASTM E 1137 Grade A Classification of Tolerances. Use the coefficients below to achieve the tolerances listed in the “No Calibration” column or calculate the Grade A tolerance using the following formula: Grade A = ± [0.13 + 0.0017|t|]°C where: |t| = value of temperature (°C) without regard to sign. See ASTM E 1137. R0: Enter the R0 value provided A: 3.9083 x 10–3 B: –5.775 x 10–7 C: –4.183 x 10–12 |
型号 | 描述 |
---|---|
5606-50-X |
探头,浸入式 PRT,50 mm,-200 到 160°C † X = 终止。指定“A”(适用于 914X 的 INFO-CON)、“B”(裸线)、“D”(适用于 Tweener 温度计的 5 针 DIN)、“G”(金针)、“I”(适用于 1521 或 1522 的 INFO-CON)、“J”(香蕉插头)、“L”(微型平接线片)、“M”(微型香蕉插头)、“P”(适用于 1523 或 1524 的 INFO-CON)或“S” (平接线片)。 |
配件 | 描述 |
---|---|
2603 |
小探头提行箱 |
1924-4-10 |
PRT 校准, |
1930-4-10 |
系统对比校准,-200 至 157° |
全球多地设有办事处
恒仪总部设在河北保定,北京、德国、美国、香港设有办事处,产品采购售后服务更方便。
全球1000多家供应商合作
恒仪与全球超过1000家工业品供应商密切合作,产品更全。
源头采购 价格更低
恒仪直接与全球各品牌厂家采购,省去中间商、代理商差价,价格更低、产品质量有保障。综合利润不超5
经验丰富 服务更好
我们的团队核心成员均来自行业知名企业,经验超过10年,为您提供更好的服务。
正品保障 更放心
每一件产品均全球各品牌原厂采购,保证100%正品、保证质量、保证售后。拒绝任何假货、次品货。
空运直发 货期更快
恒仪与国际知名物流公司合作,非限制产品,只发空运,货期更快。让客户更快拿到产品。
技术雄厚 售后保障
恒仪设有专业的技术及售后团队,为您解决各类产品使用问题,及时为您处理售后问题,服务有保障。
一站式服务 客户更轻松
恒仪采购全程一站式服务,让客户轻轻松松获得国际优质工业品,不再担心各类采购问题。